Keysight N432A Programming Manual page 66

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

1
N432A Remote Operation
CONFigure examples
The following examples demonstrate how to use the CONFigure commands,
together with READ?, INITiate?, and FETCh? to make measurements.
It is advisable to read through these examples in order as they become
increasingly more detailed.
For further information on the CONFigure commands, refer to
MEASurement
Example 1 – The simplest method
The following program segments show the simplest method of querying the
upper-window and lower-window measurement results.
Using READ?
*RST
CONF1
READ1?
*RST
CONF2
READ2?
66
Commands.
Resets the N432A.
Configures the upper-window measurement – sets to a single measurement by default.
Initiates and retrieves the upper-window measurement.
Resets the N432A.
Configures the lower-window measurement – sets to a single measurement by default.
Initiates and retrieves the lower-window measurement.
Chapter
Keysight N432A Programming Guide
3,

Advertisement

Table of Contents
loading

Table of Contents