Keysight N432A Programming Manual page 248

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

5
CALibration Subsystem
CALibration Subsystem
CALibration commands are used to perform zeroing on the N432A as well as
apply the reference calibration factor for the thermistor sensor being used.
The numeric suffix of the CALibration command (CALibration1) represents
channel A.
Zeroing of the N432A is recommended:
– when a 5
– when connection to the thermistor sensor is established.
– when you change the thermistor sensor.
– every 24 hours.
– prior to measuring low-level signals.
The CAL:ZERO:AUTO command is an overlapped command, where it allows the
N432A to continue parsing and executing subsequent commands while it is still
executing.
Keyword
CALibration[1]
:RCFactor
:ZERO
:AUTO
248
o
C change in temperature occurs.
Parameter form
<numeric_value>
<boolean>|ONCE
Note
Keysight N432A Programming Guide
Page
page 249
page 251