MEMory Subsystem
MEMory[:METer][1]:TABLe:FREQuency
<numeric_value>{,<numeric_value>}
This command is used to enter frequency data into the currently selected table.
Any previous frequency list is cleared before the new frequency list is stored. The
frequencies must be entered in the ascending order. Entries in the frequency lists
correspond to entries in the calibration/offset factor lists.
For sensor calibration tables only, the first calibration factor entered using the
NOTE
MEMory:TABLe:GAIN command is used as the reference calibration factor.
For sensor calibration tables, the number of frequency points must be one less
than the number of calibration factor points. This is verified when the sensor
calibration table is selected using SENSe:CORRection:CSET:SELect <string>.
Ensure that the frequency points you use cover the frequency range of the signals
that you want to measure. If you measure a signal with a frequency outside the
frequency range defined in the table, then the N432A uses the highest or lowest
point in the table to calculate the calibration factor/offset.
Depending on available memory, the N432A is capable of storing 20 sensor
calibration tables and 10 frequency-dependent offset tables, each containing 80
points.