Keysight N432A Programming Manual page 481

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

INITiate[:IMMediate]:SEQuence[1]
This command sets the N432A in the wait-for-trigger state. When a trigger is
received, the measurement is taken and the result is placed in the N432A
memory. If TRIGger[1]:SOURce is set to IMMediate, the measurement begins as
soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer. Refer to
"FETCh[1]|2|3|4 Queries"
This command performs the same function as INITiate[1][:IMMediate].
NOTE
Syntax
Example
INIT:SEQ
Error message
If the N432A is not in the idle state or INITiate:CONTinuous is set to ON, error
–213,"INIT ignored" occurs.
Keysight N432A Programming Guide
INIT
:IMM
This command places the N432A in the wait-for-trigger state.
on page 165 for further information.
:SEQ