N432A Remote Operation

Editing sensor calibration tables

It is not possible to delete any of the existing calibration tables or create any
additional tables. However, the existing tables can be edited using the MEMory
subsystem. To do this, perform the following procedure:
1 Select one of the existing tables using the
2 Enter the frequency data using the
3 Enter the calibration factors using the
The following table shows how entries in the frequency lists correspond with
entries in the calibration factor lists:
The number of frequency points must be one less than the number of calibration
NOTE
factor data points. The first calibration factor data point is the reference
calibration factor, and each subsequent data point is a calibration factor in the
sensor calibration table.
86
<character_data>
command.
For information on renaming the sensor calibration tables, refer to
sensor calibration tables"
tables in the N432A, refer to
page 87.
<numeric_value>{,<numeric_value>}
MEMory[:METer][1]:TABLe:GAIN[:MAGNitude]
<numeric_value>{,<numeric_value>}
Frequency
Frequency 1
"
Frequency 80
MEMory[:METer][1]:TABLe:SELect
on page 87. For information on the currently stored
"Listing stored sensor calibration tables"
MEMory[:METer][1]:TABLe:FREQuency
command.
command.
Keysight N432A Programming Guide
"Renaming
Calibration factor
Reference calibration factor
Calibration factor 1
"
Calibration factor 80
on