Keysight N432A Programming Manual page 114

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

1
N432A Remote Operation
Procedure
1 Send a device clear message to clear the N432A output buffer.
2 Clear the event registers with the *CLS (clear status) command.
3 Enable Operation Complete using the *ESE 1 command (Standard Event
4 Send the *OPC? query and enter the result to assure synchronization.
5 Send your programming command string, and place the *OPC command as the
6 Send the *STB? (status byte) query to poll the register. This command does
7 In the GPIB mode only, you can use a serial poll to check when bit 5 (Standard
114
register).
last command.
not clear the Status Byte summary register.
Event) is set in the Status Byte summary register. You can also configure the
N432A for an SRQ interrupt by sending *SRE 32 (Status Byte enable register,
bit 5).
Keysight N432A Programming Guide