Using the lower-level commands

An alternative method in making measurements is to use the lower-level
commands to set up the resolution and measurement type as follows:
DISPlay[:WINDow[1]|2]:NUMeric[1]|2:RESolution <numeric_value>
CALCulate[1]|2|3|4:MATH[:EXPRession] <string>
CALCulate[1]|2|3|4:RELative[:MAGNitude]:AUTO <boolean>|ONCE
The advantage of using the lower-level commands over the CONFigure command
is that they give you more precise control of the N432A. As shown in
the CONFigure command presets various states in the N432A. It may be likely that
you do not want to preset these states.
Example
The following example sets the measurement type to a single measurement and
the resolution setting to 3 using the lower-level commands. The measurement is
carried out on the lower window.
ABOR1
CALC2:MATH:EXPR "(SENS1)"
DISP:WIND2:RES 3
INIT1
FETC2?
Keysight N432A Programming Guide
N432A Remote Operation
Aborts the measurement.
Sets the lower window to a single measurement.
Sets the lower-window resolution setting to 3.
Sets the channel in a wait-for-trigger state.
Triggers the lower-window measurement, then retrieves the
measurement reading.