Keysight N432A Programming Manual page 171

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

Item
expected_value
(for the expected
power level)
resolution
source list
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command
[b] When the measurement result is linear, this parameter represents the number of significant digits. When the
Example
FETC2:DIFF?
Error messages
– If the last measurement is not valid, error –230,"Data corrupt or stale" occurs.
– If the expected power level and resolution parameters are not the same as the
Keysight N432A Programming Guide
Description/Defaul t
The N432A ignores the numeric value entered for this
parameter. Any value entered is treated like DEF.
A numeric value for the resolution. If unspecified, the
current resolution setting is used. If a value is entered, it
should correspond to the current resolution setting
otherwise an error occurs.
Specifies the channels for calculating the difference
value. If unspecified and the current window setup is a
difference measurement, then this setup is used,
otherwise it defaults to A–A.
subsystems. The parameters must be entered in the specified order. If parameters are omitted, they default from
the right. The parameter DEFault is used as a placeholder. Specifying DEF leaves the parameter value unchanged.
measurement result is logarithmic, 1 to 4 represent 1, 0.1, 0.01, and 0.001 respectively.
Queries the difference measurement with relative mode off
for the lower window/upper measurement.
A measurement is valid after it has been initiated. It becomes invalid when
either a reset occurs or any measurement parameter, for example frequency, is
changed.
current expected value and resolution settings on the specified window/
measurement, error –221,"Settings conflict" occurs.
MEASurement Commands
Range of values
Sensor dependent
[a]
DEF
[b]
1 to 4
1.0, 0.1, 0.01, 0.001
[a]
DEF
(@1),(@1)
3
171