[SENSe[1]:]RSELection <character_data>
This command selects the resistance type of Rmeas or Ruser for the thermistor
sensor. When Rmeas is selected, an externally measured bridge resistance value
stored in EEPROM will be used. When Ruser is selected, you can enter a
resistance value to be used in power measurement.
Syntax
Parameter
Item
character_data
Reset condition
On reset, the resistance type is set to Rmeas.
Example
RSEL MEAS
Keysight N432A Programming Guide
:
SENS
1
Description/Defaul t
Character data representing the resistance type.
This command sets the resistance type to Rmeas.
Space
RSEL
?
SENSe Subsystem
character_data
Range of values
MEAS
USER
12
349