Keysight N432A Programming Manual page 197

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

Item
expected_value
(for the expected
power level)
resolution
source list
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command
[b] When the measurement result is linear, this parameter represents the number of significant digits. When the
Example
READ:RAT:REL?
Error messages
– INITiate:CONTinuous must be set to OFF, otherwise error –213,"INIT ignored"
– If TRIGger:SOURce is set to BUS or HOLD, error –214,"Trigger deadlock" occurs.
– If the resolution parameter is not the same as the current resolution setting on
Keysight N432A Programming Guide
Description/Defaul t
The N432A ignores the numeric value entered for this
parameter. Any value entered is treated like DEF.
A numeric value for the resolution. If unspecified, the
current resolution setting is used. If a value is entered, it
should correspond to the current resolution setting
otherwise an error occurs.
Specifies the channels for calculating the ratio value. If
unspecified and the current window setup is a ratio
measurement, then this setup is used, otherwise it
defaults to A/A.
subsystems. The parameters must be entered in the specified order. If parameters are omitted, they default from
the right. The parameter DEFault is used as a placeholder. Specifying DEF leaves the parameter value unchanged.
measurement result is logarithmic, 1 to 4 represent 1, 0.1, 0.01, and 0.001 respectively.
occurs.
the specified window/measurement, error –221,"Settings conflict" occurs.
Queries the relative ratio measurement for the upper
window/upper measurement.
MEASurement Commands
Range of values
Sensor dependent
[a]
DEF
[b]
1 to 4
1.0, 0.1, 0.01, 0.001
[a]
DEF
(@1),(@1)
3
197