Keysight N432A Programming Manual page 111

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

The condition bits are set and cleared under the following conditions:
Table 1-8
Bit
0
4
5
11
12
Keysight N432A Programming Guide
Bit change conditions for the Operation Status register set
Definition
EVENts causing bit changes
This is the summary bit for the Operation CALibrating register
CALibrating
SET: At the beginning of zeroing (CALibration:ZERO:AUTO ONCE)
CLEARED: At the end of zeroing
This is the summary bit for the Operation MEASuring register
MEASuring
SET: When the N432A is taking a measurement
CLEARED: When the measurement has completed
This is the summary bit for the Operation TRIGger register
Waiting for
SET: When the N432A enters the "wait-for-trigger" state
TRIGger
CLEARED: When the N432A enters the "idle" state
This is the summary bit for the Lower Limit Fail register
Lower Limit
SET: If a measurement is made and a window lower limit test fails
Fail
CLEARED: If a measurement is made and the lower limit test is not enabled or
the test is enabled and passes
This is the summary bit for the Upper Limit Fail register
Upper Limit
SET: If a measurement is made and a window upper limit test fails
Fail
CLEARED: If a measurement is made and the upper limit test is not enabled or
the test is enabled and passes
N432A Remote Operation