Keysight N432A Programming Manual page 450

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

15
SYSTem Subsystem
SYSTem:ERRor?
This query returns error numbers and messages from the N432A error queue.
When an error is generated by the N432A, it stores an error number and its
corresponding message in the error queue. One error is removed from the error
queue each time this query is sent. The errors are cleared in the first-in first-out
order, where the oldest errors are cleared first. To clear all errors from the error
queue, execute the *CLS command. When the error queue is empty, subsequent
SYSTem:ERRor? queries will return +0,"No error". The error queue has a
maximum capacity of 30 errors.
Syntax
SYST
Example
SYST:ERR?
Reset condition
On reset, the error queue is unaffected.
Error message
If the error queue overflows, the last error stored in the queue (the most recent
error) is replaced with –350,"Queue overflow". No additional errors are stored until
you remove errors from the queue.
450
?
:ERR
Queries the oldest error message stored in the N432A error
queue.
Keysight N432A Programming Guide