CALCulate Subsystem
CALCulate[1]|2|3|4:RELative[:MAGNitude]:AUTO <boolean>|ONCE
This command sets the reference value to be used in the relative measurement for
the specified window/measurement.
The parameter should be set to ONCE to set the reference value to be used in the
relative measurement. Specifying ONCE sets the reference value to that of the
measurement signal after any math calculations and display offsets have been
applied. After the reference value has been set, the command returns to OFF.
Setting this command parameter to ONCE turns ON the
CALCulate[1]|2|3|4:RELative:STATe command.
If 0|OFF is specified, no reference value is applied to the measurement signal.
There is no situation in which you would want to send this command with OFF. OFF
is only available because it is required for the query response.
If 1|ON is selected, it causes error –224,"Illegal parameter value" to occur.
Syntax
CALC
1
2
3
4
Example
CALC:REL:AUTO ONCE
242
:MAGN
:REL
Space
:AUTO
This command sets a reference value to be used in
the relative measurement on the upper window/
upper measurement.
Keysight N432A Programming Guide
1|ON
0|OFF
ONCE
?