Keysight N432A Programming Manual page 268

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

6
DISPlay Subsystem
Parameter
Item
numeric_value
Example
DISP:WIND2:MET:LOW 10
Reset condition
On reset, the lower scale limit is set to –70 dBm.
Query
DISPlay[:WINDow[1]|2]:METer:LOWer? [MIN|MAX]
This query returns the current setting of the analog meter lower scale limit or the
value associated with MIN or MAX. The response format is <NR3>. The units in which
the results are returned are dependent on the current settings of UNIT:POWer and
CALCulate:RELative:STATe as shown in
Query example
DISP:MET:LOW?
268
Description/Defaul t
A numeric value for the analog meter lower scale limit:
– DEF: The default value is –70 dBm
– MIN: –150 dBm
– MAX: 230 dBm
The units used are determined by the current settings of
UNIT:POWer and CALCulate:RELative:STATe as
shown in
Table
6-3.
This command sets the analog meter lower scale
limit of the lower window to 10 dBm.
Queries the lower scale limit set on the analog meter in the
upper window.
Table
6-3.
Keysight N432A Programming Guide
Range of values
–150 to 230 dBm
DEF
MIN
MAX