Usb Configuration - Keysight N432A Programming Manual

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

– Domain name – SYSTem:COMMunicate:LAN:DNAMe <character_data>
– Host name – SYSTem:COMMunicate:LAN:HNAMe <character_data>
– Restart network – SYSTem:COMMunicate:LAN:RESTart
The character_data values for the IP address, subnet mask, and default gateway
can range between 0.0.0.0 and 255.255.255.255.
For details on the individual command, refer to
page 431.
If you configure an invalid IP address or an IP address that is used by another
NOTE
device or host, an error message is generated. This error can be read using the
SYSTem:ERRor? query.
The LAN setting values are stored in non-volatile memory and are not part of the
save-recall function.

USB configuration

The USB interface requires no front panel or remote configuration.
The USB address cannot be changed; it is set at the factory and is unique for each
instrument.
– For further information on the USB configuration, refer to the N432A
NOTE
– Before connecting the USB cable, make sure that I/O software is installed on
– For more information on the Keysight IO Libraries software, refer to the
– If you have installed other I/O software, refer to the documentation that
Keysight N432A Programming Guide
Thermistor Power Meter Installation Guide.
your PC.
Connectivity Guide.
accompanies the software
N432A Remote Operation
"SYSTem Subsystem"
.