Keysight N432A Programming Manual page 384

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

13
SERVice Subsystem
SERVice:BIST:CALibrator <boolean>
This command enables or disables the reference calibrator self-test during power
up. It can be used to disable the self-test if it incorrectly indicates a failure. If a
load, for example, a sensor, is connected to the reference calibrator port, this
could cause the self-test to fail. Also, if it fails the self-test, a pop-up message will
be displayed for 5 seconds, stating "If Ref Calibrator test fails disconnect any load
attached to it and re-try test".
Syntax
Example
SERV:BIST:CAL OFF
Query
SERVice:BIST:CALibrator?
This query enters a 1 or 0 into the output buffer indicating the status of the
reference calibrator self-test:
– 1 is returned if the self-test is enabled.
– 0 is returned if the self-test is disabled.
Query example
SERV:BIST:CAL?
384
SERV
:BIST
Space
:CAL
This command disables the reference calibrator
self-test during power up.
Queries whether the reference calibrator self-test is
enabled or disabled.
0|OFF
1|ON
?
Keysight N432A Programming Guide