Keysight N432A Programming Manual page 491

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

TRIGger[:SEQuence[1]]:SOURce BUS|HOLD|IMMediate
This command configures the trigger system to respond to the specified source.
This command only selects the trigger source. Use the INITiate command to
place the N432A in the wait-for-trigger state.
This command has the same purpose as TRIGger[1]:SOURce
NOTE
BUS|HOLD|IMMediate.
Syntax
TRIG
:SEQ
Parameter
Item
source
Keysight N432A Programming Guide
1
:SOUR
Description/Defaul t
Available trigger sources:
– BUS: The trigger source is the group execute trigger (GET) bus
command or the TRIGger:IMMediate SCPI command.
– HOLD: Triggering is suspended. The only way to trigger the
N432A is to use TRIGger:IMMediate.
– IMMediate: The trigger system is always true. If
INITiate:CONTinuous is set to ON, the N432A is continually
triggering free (free run mode). If an INITiate:IMMediate
command is sent, a measurement is triggered then the N432A
returns to the idle state.
Space
BUS
HOLD
IMM
?
TRIGger Subsystem
Range of values
BUS
HOLD
IMMediate
16
491

Advertisement

Table of Contents
loading

Table of Contents