4
CALCulate Subsystem
Reset condition
On reset, <string> is set to POW:AVER.
Query
CALCulate[1]|2|3|4:FEED[1]|2?
This query returns the current input measurement mode for the specified CALC
block and feed.
Query example
CALC1:FEED2?
Error message
If the command is executed when no sensor is attached, error –241,"Hardware
missing" occurs.
218
Queries the input measurement mode for feed 2 of the
upper window/upper measurement.
Keysight N432A Programming Guide