N432A Remote Operation
Reviewing table data
To review the data stored in a sensor calibration table, use the following
commands:
MEMory[:METer][1]:TABLe:SELect <character_data>
Select a sensor calibration table of the specified name. The
MEMory[:METer][1]:TABLe:SELect? query returns the name of the currently
selected table.
MEMory[:METer][1]:TABLe:FREQuency:POINts?
Returns the number of frequency points for the currently selected table.
MEMory[:METer][1]:TABLe:FREQuency?
Returns a list of frequency points (in Hz) for the currently selected table.
MEMory[:METer][1]:TABLe:GAIN[:MAGNitude]:POINts?
Returns the number of calibration factor points (including the reference
calibration factor) for the currently selected table.
MEMory[:METer][1]:TABLe:GAIN[:MAGNitude]?
Returns a list of calibration factors for the currently selected table.
Refer to the individual command under
on page 315 for details.
Modifying table data
If you need to modify the frequency and calibration factor data stored in a sensor
calibration table, you need to resend the complete data lists.
If you have retained the original data in a program, edit the program and resend
the data.