Keysight N432A Programming Manual page 196

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

3
MEASurement Commands
READ[1]|2|3|4[:SCALar][:POWer:AC]:RATio:RELative?
[<expected_value>[,<resolution>[,<source list>]]]
This query sets the measurement type to a ratio measurement with relative mode
on for the specified window/measurement, aborts then initiates the measurement,
calculates the ratio measurement result using the new sensor data, and places
the result on the bus. The result is a ratio-based measurement and is expressed in
the unit defined by UNIT[1]|2|3|4:POWer:RATio. The relative value used is that
set by the CALCulate:RELative:MAGNitude:AUTO command.
INITiate:CONTinuous must be set to OFF, otherwise error –213,"INIT ignored"
NOTE
occurs. If TRIGger:SOURce is set to BUS, error –214,"Trigger deadlock" occurs.
Syntax
READ
1
2
3
4
Parameters
Refer to
parameters in this command.
196
:POW
:SCAL
expected_value
Space
DEF
"Optional parameters"
:RAT
:AC
,
resolution
DEF
on page 146 for additional details on the
Keysight N432A Programming Guide
:REL
?
,
source list

Advertisement

Table of Contents
loading

Table of Contents