TRIGger[1]:DELay:AUTO <boolean>
This command is used to determine whether or not there is a settling-time delay
before a measurement is made.
When this command is set to:
– ON, the N432A inserts a settling-time delay before taking the requested
– OFF, the N432A makes the measurement immediately after a trigger is
TRIGger[1]:DELay:AUTO is ignored if TRIGger[1][:IMMediate] is set to ON.
Syntax
Example
TRIG:DEL:AUTO ON
Reset condition
On reset, this command is set to ON.
Keysight N432A Programming Guide
measurement. This settling time allows the internal digital filter to be updated
with new values to produce valid and accurate measurement results. The
trigger with delay command allows settling time for the internal amplifiers and
filters. It does not allow time for sensor delay.
In cases of large power changes, the delay may not be sufficient for a complete
settling. Accurate readings can be assured by taking two successive
measurements for comparison.
received.
TRIG
1
:AUTO
:DEL
This command enables a delay on the N432A.
TRIGger Subsystem
Space
0|OFF
1|ON
?
16
483