Keysight N432A Programming Manual page 249

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

CALibration[1]:RCFactor <numeric_value>
This command is used to set the value for the reference calibration factor. The
reference calibration factor can also be set from the sensor calibration table. The
N432A uses the most recently set reference calibration factor.
Syntax
Parameter
Item
numeric_value
Example
CAL:RCF 98
Keysight N432A Programming Guide
CAL
1
Description/Defaul t
A numeric value for the reference calibration factor
– DEF: The default value is 100%
– MIN:1%
– MAX: 150%
This command enters a reference calibration factor of 98%.
Space
:RCF
?
CALibration Subsystem
numeric_value
DEF
MIN
MAX
Space
MIN
MAX
Range of values
1.0 to 150.0 PCT
DEF
MIN
MAX
5
249