Keysight N432A Programming Manual page 263

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

Parameter
Item
numeric_value
Example
DISP:WIND2:ANAL:UPP 50
Reset condition
On reset, the upper scale limit is set to 20 dBm.
Query
DISPlay:[WINDow[1]|2]:ANALog:UPPer? [MIN|MAX]
This query returns the current setting of the analog meter upper scale limit, or the
value associated with MIN or MAX. The response format is <NR3>. The units in which
the results are returned are determined by the current settings of UNIT:POWer and
CALCulate:RELative:STATe as shown in
Query example
DISP:WIND2:ANAL:UPP?
Keysight N432A Programming Guide
Description/Defaul t
A numeric value for the analog meter upper scale limit:
– DEF: The default value is 20 dBm
– MIN: –150 dBm
– MAX: 230 dBm
The units used are determined by the current settings of
UNIT:POWer and CALCulate:RELative:STATe as shown
in
Table
6-2.
This command sets the analog meter upper
scale limit of the lower window to 50 dBm.
Queries the upper scale limit set on the analog
meter in the lower window.
DISPlay Subsystem
Table
6-2.
6
Range of values
–150 to 230 dBm
DEF
MIN
MAX
263