Keysight N432A Programming Manual page 186

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

3
MEASurement Commands
READ[1]|2|3|4[:SCALar][:POWer:AC]?
[<expected_value>[,<resolution>[,<source list>]]]
This query sets the measurement type to single measurement with relative mode
off for the specified window/measurement, aborts then initiates the measurement,
calculates the measurement result, and places the result on the bus. The result is
a power-based measurement and is expressed in the unit defined by
UNIT[1]|2|3|4:POWer.
INITiate:CONTinuous must be set to OFF, otherwise error –213,"INIT ignored"
NOTE
occurs. If TRIGger:SOURce is set to BUS, error –214,"Trigger deadlock" occurs.
Syntax
READ
1
2
3
4
186
:POW
:SCAL
expected_value
Space
DEF
?
:AC
,
resolution
DEF
Keysight N432A Programming Guide
,
source list

Advertisement

Table of Contents
loading