Keysight N432A Programming Manual page 233

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

CALCulate[1]|2|3|4:LIMit:UPPer[:DATA] <numeric_value>
This command enters a value for the upper test limit for the specified window/
measurement used in the CALCulate[1]|2|3|4:LIMit:FAIL? test. The units used
are dependent on the current settings of
UNIT:POWer and CALCulate:RELative:STATe as shown in
measured power is greater than the value specified in
CALCulate[1]|2|3|4:LIMit:UPPer[:DATA], CALCulate[1]|2|3|4:LIMit:FAIL?
reports a fail. When the measured power is less than or equal to the limit, a fail is
not reported.
Table 4-2
Measurement mode
Single
Ratio
Difference
Syntax
CALC
1
2
3
4
Keysight N432A Programming Guide
Measurement units
:UPP
:LIM
CALC:REL:STAT OFF
Linear
Log
Watt
dBm
%
dB
%
dB
Watt
dBm
%
dB
Space
:DATA
?
CALCulate Subsystem
Table
4-2. When the
CALC:REL:STAT ON
Linear
%
%
%
%
%
numeric_value
DEF
MIN
MAX
Space
MIN
MAX
4
Log
dB
dB
dB
dB
dB
233