Table 2-1 Zero Test Results - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

2
Performance Tests
When switching the range calibrator to STANDBY, allow enough time for the range
N O TE
calibrator to settle to its zero value before attempting to zero the power meter. This
settling would appear on the power meter display as downward drift. When the drift
has reached minimum, (typically less than 60 seconds), the range calibrator is settled.
6 Set the filter length:
• Press
• Highlight the Filter parameter using the up/down arrow keys.
• Scroll through the Filter values by pressing
MAN.
• Highlight the filter length using the forward arrow key. Press
The Filter Length pop- up window is displayed. Set the filter length to
512.
• Press
7 Zero the power meter:
Press
wait symbol to disappear.
8 Wait 30 seconds then verify that the power meter's reading is within
±0.05 μW. Record the reading.

Table 2-1 Zero Test Results

–76.40 nW
20
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Channel
.
Enter
.
Zero
,
Zero
. Wait approximately 10 seconds for the
Cal
Min
Channel A Actual
Results
_________________
. Select a value of
Change
Channel B Actual
Results (E4417A Only)
_________________
E4416A/E4417A Power Meters Service Guide
Change
.
Max
+76.40 nW