Read the Limit Test Results
The
command in line 450 returns the limit test result for failed points. The test
results are in the following order: sweep parameter, result (0 for fail, 1 for no test), upper
limit, and lower limit.
The
command in line 470 returns the number of failed points. (When the limit test
result is PASS, it returns 0 and the program goes to
as many lines as the transferred data.
Lines 510 to 560 print the limit test result as follows:
d
What are other
commands used to
retrieve the test
results?
8-4
Programming Limit T est from Remote
Instead of reading the limit test results for failed points by
using the
result using the following commands:
At all measurement points:
At marker position:
Both commands return the sweep parameter, result (1 for
pass, 0 for fail, 1 for no test), upper limit, and lower limit.
.) Then the array
command, you can read out the test
is printed with
a