3. Select the Segment Sweep as the Sweep Type

The segment sweep is selected as the sweep type.
Setup Description
Sweep type: Segment sweep

4. Execute the Calibration

In this step, a 2-port ECal module and 3-/4-port module installed in the E5070B/E5071B
are used to execute calibration on the three ports used in the measurement.
Step 1. Connect the USB port of the 2-port ECal module and the USB port of the E5070B/E5071B
with a USB cable. The connection may be made while the unit is powered.
Step 2. Load and execute the 3-/4-port ECal programs.
Setup Description
Opening the VBA Project Open dialog box
Loading "ECalAssistant.VBA"
Executing the program
Step 3. The EcalAssistant (start) dialog box appears (Figure 15-8).
Figure 15-8
EcalAssistant (start) dialog box
Step 4. Press the
(Figure 15-9).
Chapter 15
button to display the EcalAssistant (port/channel selection) dialog box
Next
Measurement Examples
Evaluating a Duplexer
Key Operation
(Moves the focus to the softkey menu)
-
-
-
Return
Sweep Type
Segment
(or
-
Sweep Type
Key Operation
-
Load Project
Select "D:\Agilent\ECalAssistant.VBA" and
press the
button.
Open
-
)
Segment
495