Agilent Technologies ENA Series User Manual page 250

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Making Measurements
Setting Up the Trigger and Making Measurements
Step 3. Press the softkey that corresponds to the desired trigger source.
Softkey
Internal
External
Manual
Bus
2. Selecting a Trigger Mode
Follow the procedure below to select a trigger mode.
Step 1. Press
be set.
Step 2. Press
Step 3. Press the softkey that corresponds to the desired trigger mode.
Softkey
Hold
Single
Continuous
Hold All Channels
Continuous Disp Channels
Step 4. Repeat Step 1 to Step 3 until each channel is set for its trigger mode.
3. Generating the Trigger
Next, it is necessary to generate a trigger by using the trigger source selected in "1.
Selecting a Trigger Source" on page 243.
NOTE
Once the internal trigger source is selected, a series of triggers is continuously generated as
soon as the setting becomes effective.
Pressing
244
Function
Selects internal trigger source
Selects external trigger source
Selects manual trigger source
Selects bus trigger source
(or
) to select the channel for which the trigger mode will
.
Function
Sets active channel trigger mode to hold sweep mode
Sets active channel trigger mode to single sweep mode
Sets active channel trigger mode to continuous sweep mode
Sets all channel trigger modes to hold sweep mode
Sets trigger modes of all displayed channels (
to continuous sweep mode
-
during a sweep forces the analyzer to abort the sweep.
Restart
-
)
Display
Chapter 5

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5070bE5071b

Table of Contents