Step 7. In the same way, repeat Step 2 to Step 6 to enter the definition of REFLECT to No. 2.Select
SHORT for STD Type.
Procedure to define Match
Step 1. Press
-
3:No Name
Step 2. Type in
MATCH <2G
Step 3. Press
-
STD Type
Load
Step 4. Set
Max Frequency
Step 5. Press
to return to the Define Std menu.
Return
Procedure to define Line 1/2
Step 1. Press
-
4:No Name
Step 2. Type in
.
LINE <7G
Step 3. Press
STD Type - Thru
Step 4. Set
to 54ps.
Offset Delay
Step 5. Set
Min Frequency
Step 6. Set
Max Frequency
Step 7. Press
to return to the Define Std menu.
Return
Step 8. Press
-
5:No Name
Step 9. Type in
.
LINE >7G
Step 10. Press
-
STD Type
Thru
Step 11. Set
to 13ps.
Offset Delay
Step 12. Set
Min Frequency
Step 13. Press
to return to the Define Std menu.
Return
Step 14. Press
to return to the Modify Cal Kit menu.
Return
Procedure to register Thru/Reflect/Match to sub class 1
Step 1. Press
Specify CLSs
Step 2. Select
.
Sub Class1
Step 3. Press
TRL Thru - Set All - THRU - Return
Step 4. Press
TRL Reflect - REFLECT
Step 5. Press
TRL Line/Match - Set All - MATCH <2G - Return
Procedure to register Line 1/2 to subclass 2/3
Step 1. Select
.
Sub Class2
Step 2. Press
TRL Line/Match - Set All - LINE <7G - Return
Chapter 4
Changing the Calibration Kit Definition
.
Label
.
.
to 2GHz.
.
Label
.
to 2GHz.
to 7GHz.
.
Label
.
to 7GHz.
-
.
Sub Class
.
.
Calibration
.
.
189