Comparing Functions - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Comparing Functions

Table F-5
Functions of the 8753ES vs. the E5070B/E5071B
Function
Measurement Reset
Channel
Number of channels
Coupling between channels
Trace
Measurement parameter
S-parameter conversion
Display format (Data format)
Test port output Number of ports
Frequency
Power level
Characteristic Impedance
Coupled/Uncoupled power
levels
Power slope function
Turning the output On/Off
Power range
Appendix F
Replacing the 8753ES with the E5070B/E5071B
8753ES
Can be executed by using the front
panel and the GPIB command.
4 channels
Channels 3 and 4 are auxiliary
channels for channels 1 and 2 and
subject to the same sweep
conditions, etc. applicable to
channels 1 and 2.
Couple/Uncouple between channels
1 and 2 can be set up.
A trace for each channel
S-parameters, A, B, R, A/R, B/R,
A/B, and analog bus
Impedance (reflection and
transmission), admittance (reflection
and transmission), and 1/S
log magnitude, linear magnitude,
phase, group delay, Smith chart,
polar format, SWR, real, imaginary
2 ports
30 kHz to 3 GHz (Std.)
30 kHz to 6 GHz (Opt. 006)
-85 to +10 dBm (Std.)
-85 to +8 dBm (Opts. 014/075)
50 Ω(Std.)
75 Ω(Opt. 075)
You can define Couple/Uncouple
between channels and between ports.
Available
Allowed
Items to be set
Per port and per channel
up
Automatic
On/Off setting capability
setting
Setting range
Range setting with 10 dBm
resolution is definable from between
-15 and +10 dBm down to between
-85 and -60 dBm.
Comparing Functions
E5070B/E5071B
Can be executed by using the front
panel, the GPIB command, and
telnet.
4/9/12/16 channels
Channels are independent of each
other.
Each channel can accommodate up
to 4/9/12/16 traces.
S-parameters (Mixed-mode
S-parameters),
Absolute value meas parameter
log magnitude, linear magnitude,
phase, expanded phase, positive
phase, group delay, Smith chart,
polar format, SWR, real, imaginary
2 ports (Opts. 213/214)
3 ports (Opts. 313/314)
4 ports (Opts. 413/414)
300 kHz to 3 GHz (E5070B)
300 kHz to 8.5 GHz (E5071B)
-15 to 10 dBm (Opts. 213, 313, 413)
-50 to 10 dBm (Opts. 214, 314, 414)
50 Ω
Traces are coupled on the same
channel, not between channels.
Per channel (Opts. 214/314/414)
If the above options are not installed,
the default range is -15 to 10 dBm.
Manual setting
Range setting with 5 dBm resolution
is definable from between -15 and 10
dBm down to between -50 and -25
dBm
745