Executing Power Calibration; Overview Of Execution Procedure; Connecting Power Meter - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Executing Power Calibration

This section shows an example of executing power calibration using the E4418B power
meter and the E4412A power sensor.

Overview of execution procedure

In this example, power calibration is executed according to the steps shown in Table 15-11.
Table 15-11
Execution Procedure of Power Calibration
Step
"1. Connecting Power Meter" on page 529
"2. Setting Address of Power Meter" on
page 530
"3. Setting Stimulus Condition" on page 530
"4. Executing Zero Adjustment and
Calibration of Power Sensor" on page 530
"5. Setting Calibration Data Measurement
Conditions" on page 530
"6. Connecting Power Sensor" on page 531
"7. Measuring Calibration Data" on page 531

1. Connecting Power Meter

Connect the E4418B (GPIB address: 14) to the E5070B/E5071B as shown in Figure 15-39.
Figure 15-39
Connecting power meter
Chapter 15
Executing Power Calibration
Description
Connects the power meter to the E5070B/E5071B.
Configures the power meter's GPIB address with the
E5070B/E5071B.
Sets conditions such as the power level and frequency.
Executes zero adjustment and calibration of the power
sensor.
Selects the port, selects the power sensor, sets the number
of measurements at one point, and sets the
.
power calibration
Connects the power sensor.
Executes the measurement of calibration data.
Measurement Examples
tolerance during
529