Specifications and Supplemental Information
Measurement Throughput Summary
Table 16-22
Typical Cycle Time for Measurement Completion
Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
*1.Typical performance.
*2.Sweep mode: Std Stepped. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1. System
error correction: ON
Table 16-23
Cycle Time (ms)
Number of
Sweep mode: Fast Swept
Points
System error correction: OFF
3
4
11
4
51
4
101
4
201
5
401
8
801
11
1601
18
*1.Typical performance.
*2.Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth, Error correction: OFF, Display update: OFF, Number of traces
= 1.
556
Number of Points
51
201
7
17
12
32
14
27
26
50
16
30
30
57
*1*2
vs. Number of Points
Sweep mode: Fast Swept
System error correction: ON
4
4
4
5
6
8
13
23
*1*2
(ms)
401
1601
29
90
55
178
43
130
84
258
49
146
96
291
Sweep mode: Std Stepped
System error correction: ON
4
4
7
11
17
29
52
90