Measurement Examples
Measuring the SAW Bandpass Filter Using the Segment Sweep
Measuring the SAW Bandpass Filter Using the Segment
Sweep
This section illustrates how to use the segment sweep function to evaluate a SAW bandpass
filter with a center frequency of 947.5 MHz.
Evaluation Steps
Here, the DUT is evaluated by following the steps described in Table 15-1.
Table 15-1
Evaluating the DUT Using the Segment Sweep
Step
page 490
page 491
page 491
486
Description
The segment sweep conditions are determined according to the
characteristics of the DUT.
The segment sweep conditions are entered in the E5070B/E5071B.
The segment sweep is selected as the sweep type.
A 2-port ECal is performed between the test ports connecting the
DUT.
The DUT is connected.
A trigger is applied to perform the measurement.
The choice is made between frequency base and order base as the
method of displaying segments.
Chapter 15