Agilent Technologies ENA Series User Manual page 672

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Softkey Functions
Calibration Menu
Key Operation
(Continued)
Calibrate
Displays softkeys for executing 1-port calibrations.
1-Port Cal
Displays softkeys for selecting a test port.
Select Port
Selects test port 1.
1
Selects test port 2.
2
Selects test port 3.
*1
3
*2
Selects test port 4.
4
Select Port
Returns to the softkey display screen one level higher.
Cancel
Executes an OPEN calibration on the test port selected in the
Open
menu.
Port
Executes a SHORT calibration on the test port selected in the
Short
menu.
Port
Executes a LOAD calibration on the test port selected in the
Load
menu.
Port
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels a calibration process and clear the measurement data of
OK
standard.
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing a full 2-port calibration.
2-Port Cal
Displays softkeys for selecting test ports.
*1
Select Ports
Selects test ports 1 and 2.
1-2
Selects test ports 1 and 3.
1-3
*2
Selects test ports 1 and 4.
1-4
Selects test ports 2 and 3.
2-3
*2
Selects test ports 2 and 4.
2-4
Selects test ports 3 and 4.
*2
3-4
Returns to the softkey display screen one level higher.
Cancel
Displays softkeys for executing reflection calibrations. Symbols x and y refer to the two test ports selected in the
Reflection
Select Ports
Executes an OPEN reflection calibration on test port x.
Port x Open
Executes a SHORT reflection calibration on test port x.
Port x Short
Executes a LOAD reflection calibration on test port x.
Port x Load
Executes an OPEN reflection calibration on test port y.
Port y Open
Executes a SHORT reflection calibration on test port y.
Port y Short
Executes a LOAD reflection calibration on test port y.
Port y Load
Returns to the softkey display screen one level higher.
Return
666
Function
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
SCPI Command
:SENS{1-16}:CORR:COLL:METH:SOLT1 1
:SENS{1-16}:CORR:COLL:METH:SOLT1 2
:SENS{1-16}:CORR:COLL:METH:SOLT1 3
:SENS{1-16}:CORR:COLL:METH:SOLT1 4
:SENS{1-16}:CORR:COLL:OPEN
Select
:SENS{1-16}:CORR:COLL:SHOR
Select
:SENS{1-16}:CORR:COLL:LOAD
Select
:SENS{1-16}:CORR:COLL:SAVE
:SENS{1-16}:CORR:COLL:CLE
:SENS{1-16}:CORR:COLL:METH:SOLT2 1,2
:SENS{1-16}:CORR:COLL:METH:SOLT2 1,3
:SENS{1-16}:CORR:COLL:METH:SOLT2 1,4
:SENS{1-16}:CORR:COLL:METH:SOLT2 2,3
:SENS{1-16}:CORR:COLL:METH:SOLT2 2,4
:SENS{1-16}:CORR:COLL:METH:SOLT2 3,4
:SENS{1-16}:CORR:COLL:OPEN x
:SENS{1-16}:CORR:COLL:SHOR x
:SENS{1-16}:CORR:COLL:LOAD x
:SENS{1-16}:CORR:COLL:OPEN y
:SENS{1-16}:CORR:COLL:SHOR y
:SENS{1-16}:CORR:COLL:LOAD y