Softkey
Function
Executes line/match calibration for test ports 2 and 4.Use this to perform
2-4 Fwd (S42)
forward-direction measurement only.
Executes line/match calibration for test ports 2 and 4.Use this to perform
2-4 Rvs (S24)
reverse-direction measurement only.
Executes line/match calibration for test ports 3 and 4. Use this to perform
3-4 Line/Match
forward-direction measurement and reverse-direction measurement at the
same time.
Executes line/match calibration for test ports 3 and 4. Use this to perform
3-4 Fwd (S43)
forward-direction measurement only.
Executes line/match calibration for test ports 3 and 4. Use this to perform
3-4 Rvs (S34)
reverse-direction measurement only.
*1.This is equivalent to measuring
*2.Measuring
1-2 Fwd (S21)
Step 9. Press
to finish TRL 4-port calibration. At this point, the calibration coefficient is
Done
calculated and saved. The error correction function is automatically turned on.
Chapter 4
and
1-2 Fwd (S21)
1-2 Rvs (S12)
and
is equivalent to measuring
1-2 Rvs (S12)
Calibration
4-port TRL calibration
.
.
1-2 Line/Match
147