Key Operation
(Continued)
Calibrate
Adapter Removal
Displays softkeys for executing adapter removal/insertion.
Displays softkeys for selecting a test port.
Select Port
Selects test port 1.
1
Selects test port 2.
2
*1
Selects test port 3.
3
Selects test port 4.
*2
4
Returns to the softkey display screen one level higher.
Cancel
Displays softkeys for selecting calibration kits.
Cal Kit
Executes an OPEN calibration on the test port selected in the
Open
menu.
Port
Executes a SHORT calibration on the test port selected in the
Short
menu.
Port
Executes a LOAD calibration on the test port selected in the
Load
menu.
Port
Sets the adapter length on the test port selected in the
Adapter Length
menu.
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels a calibration process and clears the measurement data of
OK
standard.
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Appendix D
Function
Select
Select
Select
Select Port
Softkey Functions
Calibration Menu
SCPI Command
:SENS{1-16}:CORR:COLL:METH:ADAP:REM
1
:SENS{1-16}:CORR:COLL:METH:ADAP:REM
2
:SENS{1-16}:CORR:COLL:METH:ADAP:REM
3
:SENS{1-16}:CORR:COLL:METH:ADAP:REM
4
:SENS{1-16}:CORR:COLL:CKIT
:SENS{1-16}:CORR:COLL:OPEN
:SENS{1-16}:CORR:COLL:SHOR
:SENS{1-16}:CORR:COLL:LOAD
:SENS{1-16}:CORR:COLL:ADAP{1-4}:LENG
:SENS{1-16}:CORR:COLL:SAVE
:SENS{1-16}:CORR:COLL:CLE
673