4. Setting Calibration Kit
Select a calibration kit for each test port and each pair of test ports and select a standard to
use for each reflection/transmission measurement.
Step 1. Press
The Specify Cal Kit dialog box (Reflection tab) shown in Figure 4-37 appears.
Figure 4-37
Specify Cal Kit dialog box (Reflection tab)
Step 2. For each test port, select a calibration kit (1 in Figure 4-37) and a standard to use in the
open/short/load reflection measurements from the calibration kits (2 in Figure 4-37).
Step 3. Select the
Figure 4-38 appears.
Figure 4-38
Specify Cal Kit dialog box (Thru tab)
Step 4. For each test port, select a calibration kit (1 in Figure 4-38). In addition to the 10
calibration kits you can select for normal calibration, you can also select Adapter File.
When you want to use the adapter as the standard, select Adapter File.
From the calibration kits you have selected, select a standard to use for transmission
measurement (2 in Figure 4-38). If you select Adapter File as the calibration kit, you can
select an adapter file (2-port Touchstone file) under the
D:\Agilent\Data\AdapterCharacterization folder as the standard. In this case, you have to
specify the port connection mode (FWD/RVS) between the E5070B/E5071B and the
Chapter 4
Calibration between Ports of Different Connector Types
(2 in Figure 4-36).
Specify Cal Kit
tab (3 in Figure 4-37). The Specify Cal Kit dialog box (Thru tab) shown in
Thru
Calibration
157