Agilent Technologies ENA Series User Manual page 599

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Calibration Type
3-port TRL/LRM cal. for
Ports: 1-2-3
3-port TRL/LRM cal. for
Ports: 1-2-4
3-port TRL/LRM cal. for
Ports: 1-3-4
3-port TRL/LRM cal. for
Ports: 2-3-4
4-port TRL/LRM cal. for
Ports: 1-2-3-4
Step 1. Measure calibration data.
In this example, 3-port TRL/LRM calibration for test ports 1, 2, and 3 (1 in Figure A-8) is
performed by using the Match, Line 1, and Line 2 standards.
Figure A-8
Performing multiport TRL/LRM calibration
Step 2. Press the
Step 3. Make the through condition for test ports 1 and 2 and press the
Figure A-9). The through measurement is performed, and an asterisk appears at the
button.
1-2
Step 4. Make the through condition for test ports 1 and 3, and press the
Figure A-9). The through measurement is performed, and an asterisk appears at the
button and then a check mark [v] (3 in Figure A-9) also appears on the
1-3
Appendix A
Measurement ports of
Reflection and Match cal.
All ports (1, 2, and 3)
All ports (1, 2, and 4)
All ports (1, 3, and 4)
All ports (2, 3, and 4)
All ports (1, 2, 3, and 4)
button (2 in Figure A-8).
Thru
Manual Changes
Manual Changes
Measurement paths of Thru
and Line cal.
1-2 and 1-3
1-2 and 1-4
1-3 and 3-4
2-3 and 3-4
1-2, 1-3, and 3-4
button (1 in
[ ]Thru 1-2
[ ]Thru
button (2 in
[ ]Thru 1-3
[ ]Thru
button.
Thru
593