Softkey Functions
Calibration Menu
Key Operation
(Continued)
Calibrate
2-Port TRL Cal
Displays softkeys for executing THRU/LINE calibrations. Symbols x and y refer to the two test ports selected in
Thru/Line
the
Select Ports
Executes a THRU/LINE calibration on test ports x and y.
x-y Thru/Line
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing reflection calibrations. Symbols x and y refer to the two test ports selected in the
Reflect
Select Ports
Executes a OPEN/SHORT reflection calibration on test port x.
Port x
Reflect
Executes a OPEN/SHORT reflection calibration on test port y.
Port y
Reflect
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing LINE/MATCH calibrations. Symbols x and y refer to the two test ports selected in
Line/Match
the
Select Ports
Executes a LINE/MATCH calibration on test ports x and y.
x-y
Line/Match
Executes a forward LINE/MATCH calibration on test ports x and y.
x-y Fwd(Syx)
Executes a reverse LINE/MATCH calibration on test ports x and y.
x-y Rvs(Sxy)
Returns to the softkey display screen one level higher.
Return
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels a calibration process and clears the measurement data of
OK
standard.
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing 3-port TRL calibrations.
*1
3-Port TRL Cal
Displays softkeys for selecting test ports.
*2
Select Ports
Selects test ports 1, 2, and 3.
1-2-3
Selects test ports 1, 2, and 4.
1-2-4
Selects test ports 1, 3, and 4.
1-3-4
Selects test ports 2, 3, and 4.
2-3-4
Returns to the softkey display screen one level higher.
Cancel
Displays softkeys for executing THRU/LINE calibrations. Symbols x, y and z refer to the three test ports selected
Thru/Line
in the
Select Ports
Executes a THRU/LINE calibration on test ports x and y.
x-y Thru/Line
Executes a THRU/LINE calibration on test ports x and z.
x-z Thru/Line
Executes a THRU/LINE calibration on test ports y and z.
y-z Thru/Line
Returns to the softkey display screen one level higher
Return
670
Function
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
SCPI Command
:SENS{1-16}:CORR:COLL:TRLT x,y
:SENS{1-16}:CORR:COLL:TRLT y,x
:SENS{1-16}:CORR:COLL:TRLR x
:SENS{1-16}:CORR:COLL:TRLR y
:SENS{1-16}:CORR:COLL:TRLL x,y
:SENS{1-16}:CORR:COLL:TRLL y,x
:SENS{1-16}:CORR:COLL:TRLL y,x
:SENS{1-16}:CORR:COLL:TRLL x,y
:SENS{1-16}:CORR:COLL:SAVE
:SENS{1-16}:CORR:COLL:CLE
:SENS{1-16}:CORR:COLL:METH:TRL3 1,2,3
:SENS{1-16}:CORR:COLL:METH:TRL3 1,2,4
:SENS{1-16}:CORR:COLL:METH:TRL3 1,3,4
:SENS{1-16}:CORR:COLL:METH:TRL3 2,3,4
:SENS{1-16}:CORR:COLL:TRLT x,y
:SENS{1-16}:CORR:COLL:TRLT y,x
:SENS{1-16}:CORR:COLL:TRLT x,z
:SENS{1-16}:CORR:COLL:TRLT z,x
:SENS{1-16}:CORR:COLL:TRLT y,z
:SENS{1-16}:CORR:COLL:TRLT z,y
Appendix D