The line measurement is performed, and an asterisk appears at the
Step 18. Connect the Line 2 standard between test ports 1 and 3 and press the
The line measurement is performed, and an asterisk appears at the
then a check mark [v] also appears on the
NOTE
When the frequency ranges of the line and the match overlap, the data of the line is used in
the overlapped frequency area. When the frequency ranges of multiple lines overlap, the
data of the line measured later is used in the overlapped frequency areas.
NOTE
Isolation calibration measurement is not available.
Step 19. When all of the necessary calibration measurements are completed, press the
Coef
correction is turned ON. In the calibration property display (ON/OFF with
Property
4. Closing macro Press the

Change11

With the firmware version A.06.00 or lower, follow the steps below to save measurement
data into a file in Touchstone format.
Saving data in Touchstone format
Use the following VBA macro to save measurement data into a file in Touchstone format.
Folder
D:\Agilent
NOTE
Don't delete this VBA macro. This VBA macro cannot be restored by executing system
recovery.
This VBA macro saves measurement data of any channel into a Touchstone format file,
based on 1- to 4-port models.
NOTE
You can save data in "real number - imaginary number", "dB - angle" or "amplitude -
angle."
You can use data saved in Touchstone format for a circuit simulator such as Agilent
Advanced Design System (ADS) on your PC (personal computer) or workstation. For
more information on the ADS, refer to the operation manual that comes with the system.
NOTE
You cannot recall data saved in Touchstone format on the E5070B/E5071B.
For information on data structure in a saved file, refer to "Data structure in Touchstone
file" on page 597.
Appendix A
button (8 in Figure A-9). The calibration coefficient is calculated and the error
), the state is indicated by "F" (same as full n-port SOLT calibration).
button (9 in Figure A-9) to exit from the macro.
X
VBA macro name (project name)
SaveToTouchstone.vba
Manual Changes
[ ]Line2 1-2
[ ]Line2 1-3
[ ]Line2 1-3
button.
Line2
Manual Changes
button.
button.
button and
Update Cal
-
[Cal]
595