Measurement Examples

Measuring Parameters with Cable

Measuring Parameters with Cable
This section introduces an example of how to detect the location of a mismatch that occurs
in a cable by using the time domain function.

Overview of evaluation procedure

In this example, a DUT is evaluated according to the steps shown in Table 15-8.
Table 15-8
Evaluation Procedure for Deviation from Linear Phase
Step
"1. Setting the Measurement
Conditions" on page 518
"2. Executing Calibration" on
page 518
"3. Connecting the DUT" on page 519
"4. Auto Scale" on page 519
"5. Setting the Time Domain
Function" on page 520

1. Setting the Measurement Conditions

Follow these steps to set the measurement conditions:
Setting Description
Presetting
Stop frequency: 3 GHz
Number of points: 201
Specifying the low-pass mode sweep condition
Measurement parameter: S11

2. Executing Calibration

By following "1-Port Calibration (reflection test)" on page 113, execute 1-port calibration
on port 1.
518
Description
Set the measurement conditions.
Execute calibration.
Connect the DUT.
Execute auto scale.
Set the time domain function.
Key Operation
- Set Freq Low Pass
- OK
- Points -
- Transform
- S11
Chapter 15