Advertisement

NOTE
When you define each calibration standard, the following points should be considered.
Reference impedance Z0 must be set to the same value as the ENA's system impedance
Z0 value.
When you use Line 1, Line 2, and Line 3, their defined frequency ranges must overlap
by at least 10 kHz.
Saving and loading calibration kit definitions
Press the
to a file. Press the
kit from a previously saved file.
NOTE
If you save as "D:\Agilent\Data\TRL_LRM_cal\Default.dat," the file is handled as the
default definition file. The default definition file is automatically recalled when the macro
starts.
NOTE
The factory-shipped default definition file has the same content as
"D:\Agilent\Trldata\SysDefault.dat." Copy "D:\Agilent\Trldata\SysDefault.dat" to the
default definition file in order to restore the default definition file to its factory-shipped
condition.
Do not change "D:\Agilent\Trldata\SysDefault.dat."
Initializing calibration kit definition
Press the
the default definition file ("D:
Step 3. Press the
3. Performing Calibration
Measure necessary calibration data and enable error correction.
NOTE
The definition of the frequency range of the line standard and match standard used for
measurement must cover the sweep range of the channel for which you perform
calibration.
If the
definition menu and confirm that the frequency range definitions of match and lines are
correctly entered to cover the ENA's measurement frequency range.
Step 1. Measure calibration data.
In this example, 2-port TRL/LRM calibration between test ports 1 and 2 (1 in Figure A-5)
is performed by using the Match, Line 1, and Line 2 standards.
Appendix A
button (11 in Figure A-4) to save the definition of the current calibration kit
Save
button (12 in Figure A-4) to recall the definition of a calibration
Recall
button (13 in Figure A-4) to recall the definition of a calibration kit from
default
\
Agilent
button (14 in Figure A-4) to finish defining the calibration kit.
Close
,
,
, and
Thru
Reflection
Match
\
Data\TRL_LRM_cal\Default.dat").
keys do not appear, go back to the calibration kit
Line
Manual Changes
Manual Changes
589

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5070bE5071b

Table of Contents