Agilent Technologies ENA Series User Manual page 419

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Fast mode
The fast mode is a sweep mode in which the sweep time is shortened by decreasing the
waiting time before sampling (stepped mode) or speeding up the sweep (swept mode) to
the limit of the analog performance.
You can always shorten the sweep time by using the fast stepped mode rather than the
stepped mode. On the other hand, although there is some difference depending on the IF
bandwidth setting, when the measurement point interval is approximately 2 MHz or higher,
you can shorten the sweep time by using the fast swept mode rather than the swept mode.
Note that, because the instrument is adjusted to the limit of the analog performance when
using the fast mode, a correct measurement may not be obtained due to factors such as
increased trace noise, depending on the DUT and measurement conditions. In addition, for
the fast mode, the instrument specifications are not guaranteed.
NOTE
When you use the fast mode, you should confirm that there is no measurement-related
problem before performing the actual measurement.
Procedure to select the sweep mode
Step 1. Press
Step 2. Press
Step 3. Press the softkey corresponding to the appropriate sweep mode.
Softkey
Std Stepped
Std Swept
Fast Stepped
Fast Swept
Chapter 12
.
.
Sweep Mode
Function
Selects the stepped mode
Selects the swept mode
Selects the fast stepped mode
Selects the fast swept mode
Optimizing Measurements
Improving Measurement Throughput
413