Setting Measurement Conditions
Setting Stimulus Conditions
3. Enter the power level using the ENTRY block keys on the front panel.
Correcting attenuation of power level (using power slope feature)
You can use the power slope feature to correct the attenuation of a power level so that it is
simply proportional to the frequency (attenuation due to cables and so on), which improves
the accuracy of the level actually applied to the DUT.
Turning power slope feature on/off
Step 1. Press
will be turned on or off.
Step 2. Press
Step 3. Press
.
Power
Step 4. Press
Slope [OFF]
Setting correction coefficient (correction amount for 1 GHz)
Step 1. Press
correction coefficient.
Step 2. Press
Step 3. Press
.
Power
Step 4. Press
Slope [xxx dB/GHz]
Step 5. Enter the correction coefficient using the ENTRY block keys on the front panel.
76
or
to select the channel for which the power slope feature
.
(
). Each press toggles between on/off.
Slope [ON]
or
to select the channel for which you want to set the
.
. "
" represents the current set value.
xxx
Chapter 3