Agilent Technologies ENA Series User Manual page 670

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Softkey Functions
Calibration Menu
Key Operation
(Continued)
Calibrate
Displays softkeys for selecting options for response calibrations using the THRU standard.
Response (Thru)
Displays softkeys for selecting test ports.
Select Ports
Selects the transmission test (measurement of S
2-1 (S21)
1→2.
*1
Selects the transmission test (measurement of S
3-1 (S31)
1→3.
Selects the transmission test (measurement of S
*2
4-1 (S41)
1→4.
Selects the transmission test (measurement of S
1-2 (S12)
2→1.
Selects the transmission test (measurement of S
*1
3-2 (S32)
2→3.
Selects the transmission test (measurement of S
*2
4-2 (S42)
2→4.
Selects the transmission test (measurement of S
*1
1-3 (S13)
3→1.
Selects the transmission test (measurement of S
*1
2-3 (S23)
3→2.
Selects the transmission test (measurement of S
*2
4-3 (S43)
3→4.
Selects the transmission test (measurement of S
*2
1-4 (S14)
4→1.
Selects the transmission test (measurement of S
*2
2-4 (S24)
4→2.
*2
Selects the transmission test (measurement of S
3-4 (S34)
4→3.
Returns to the softkey display screen one level higher.
Cancel
Executes a THRU response calibration on the test ports selected in
Thru
the
Select Ports
transmission tracking errors.
Executes an isolation calibration on the test ports selected in the
Isolation
(Optional)
Select Ports
isolation errors.
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels a calibration process and clears the measurement data of
OK
standard.
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing enhanced response calibrations.
Enhanced
Response
Displays softkeys for selecting a test port.
Select Port
Selects the transmission test (measurement of S
2-1 (S21 S11)
1→2.
664
Function
menu. This calibration is effective for eliminating
menu. This calibration is effective for eliminating
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 2,1
21
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 3,1
31
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 4,1
41
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 1,2
12
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 3,2
32
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 4,2
42
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 1,3
13
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 2,3
23
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 4,3
43
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 1,4
14
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 2,4
24
) for test port
:SENS{1-16}:CORR:COLL:METH:THRU 3,4
34
:SENS{1-16}:CORR:COLL:THRU
:SENS{1-16}:CORR:COLL:ISOL
:SENS{1-16}:CORR:COLL:SAVE
:SENS{1-16}:CORR:COLL:CLE
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 2,1
21
SCPI Command
Appendix D

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5070bE5071b

Table of Contents