Data Output
Saving Trace Data to a File
Figure 10-7
4-port Touchstone file
Restrictions when saving data in Touchstone format
The following restrictions apply when saving measurement data into Touchstone format.
Saving procedure
Follow the steps below to save trace data in Touchstone format.
Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press the softkey that corresponds to the data format you want to save.
Softkey
Auto
364
When both fixture simulation and port impedance conversion are on, all Z0 of the ports
to be saved must be set to the same value. If Z0 is different among the ports, no error
occurs, but only the Z0 of the smallest port number is output to the header.
When the time domain function is on, the saved data are not the displayed data but the
data of the S parameter before conversion.
For data saved in touchstone format, data operation, time domain, parameter
conversion, data format, and smoothing are not reflected in the output data.
An error occurs when attempting to save data that use the frequency offset function.
or
.
-
.
SnP
SnP Format
Function
Data format is automatically set according to the display format of
the active trace.
to activate a channel in which a trace exists to be saved.
*1
Chapter 10