Agilent Technologies ENA Series User Manual page 123

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Softkey display
1-4(S14 S44)
2-4(S24 S44)
3-4(S34 S44)
NOTE
An asterisk (*) in the upper right of the softkey indicates that, if you select the test port and
execute acquisition of the calibration coefficient (press
coefficient will be cleared.
Step 6. Connect an OPEN calibration standard to the output port.
Step 7. Press
Step 8. Disconnect the OPEN calibration standard that was connected in Step 8 and replace it with
a SHORT calibration standard.
Step 9. Press
Step 10. Disconnect the SHORT calibration standard that was connected in Step 8 and replace it
with a LOAD standard.
Step 11. Press
Step 12. Make a THRU connection between the two ports.
Step 13. Press
Step 14. If an isolation calibration must be performed using a LOAD standard, follow the procedure
below.
a. Connect a LOAD standard to the two test ports.
b. Press
Step 15. Press
Step 16. Press
calibration coefficients will be calculated and saved. The error correction function will also
be automatically enabled.
Chapter 4
Function
Measure between test ports 4 and 1. selecting port 4 as the output.
Measure between test ports 4 and 2. selecting port 4 as the output.
Measure between test ports 4 and 3. selecting port 4 as the output.
to start the calibration measurement.
Open
to start the calibration measurement.
Short
to start the calibration measurement.
Load
to start the calibration measurement.
Thru
to start the calibration measurement.
Isolation (Optional)
.
Return
to terminate the enhanced response calibration process. Upon pressing the key,
Done
Calibration
Enhanced Response Calibration
), the existing calibration
Done
117