Measuring the Deviation from a Linear Phase

This section illustrates how to determine the deviation from a linear phase in the pass band
of a 1.09-GHz bandpass filter.

Evaluation Steps

Here, the DUT is evaluated by following the steps described in Table 15-5.
Table 15-5
Evaluating the Deviation from a Linear Phase
Step
"1. Connect the DUT" on page 505
"2. Define the Measurement Conditions" on
page 506
"3. Execute the Calibration" on page 506
"4. Connect the DUT and Execute the Auto
Scale" on page 506
"5. Specify the Electrical Delay" on page 507
"6. Measure the Deviation from a Linear
Phase" on page 508

1. Connect the DUT

Connect the DUT as shown in Figure 15-21.
Figure 15-21
Connecting the DUT
Chapter 15
Measuring the Deviation from a Linear Phase
Description
The DUT is connected.
The measurement conditions are defined.
The calibration is executed.
The DUT is connected again to execute the auto scale
function.
The electrical delay is specified.
The statistics data function (peak-to-peak) is used to
determine the deviation from a linear phase.
Measurement Examples
505