Calibration Menu - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Calibration Menu

Key Operation
Displays softkeys for setting and executing calibrations.
Enables or disables error correction.
Correction
Displays softkeys for selecting calibration options.
Calibrate
Displays softkeys for selecting options for response calibration using the OPEN standard.
Response (Open)
Displays softkeys for selecting a test port.
Select Port
Selects test port 1.
1
Selects test port 2.
2
Selects test port 3.
*1
3
*2
Selects test port 4.
4
Returns to the softkey display screen one level higher.
Cancel
Executes a response calibration using the OPEN standard on the
Open
test port selected in the
effective for eliminating response tracking errors.
Executes an isolation calibration using the LOAD standard on the
Load (Optional)
test port selected in the
effective for eliminating directivity errors.
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels a calibration process and clears the measurement data of
OK
standard.
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Displays softkeys for selecting options for a response calibration using the SHORT standard.
Response (Short)
Displays softkeys for selecting a test port.
Select Port
Selects test port 1.
1
Selects test port 2.
2
Selects test port 3.
*1
3
*2
Selects test port 4.
4
Returns to the softkey display screen one level higher.
Cancel
Executes a SHORT calibration on the test port selected in the
Short
menu. This calibration is effective for eliminating reflection
Port
tracking errors.
Executes an isolation calibration using the LOAD standard on the
Load (Optional)
test port selected in the
effective for eliminating directivity errors.
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels a calibration process and clears the measurement data of
OK
standard.
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Appendix D
Function
menu. This calibration is
Select Port
menu. This calibration is
Select Port
menu. This calibration is
Select Port
Softkey Functions
Calibration Menu
SCPI Command
:SENS{1-16}:CORR:STAT
:SENS{1-16}:CORR:COLL:METH:OPEN 1
:SENS{1-16}:CORR:COLL:METH:OPEN 2
:SENS{1-16}:CORR:COLL:METH:OPEN 3
:SENS{1-16}:CORR:COLL:METH:OPEN 4
:SENS{1-16}:CORR:COLL:OPEN
:SENS{1-16}:CORR:COLL:LOAD
:SENS{1-16}:CORR:COLL:SAVE
:SENS{1-16}:CORR:COLL:CLE
:SENS{1-16}:CORR:COLL:METH:SHOR 1
:SENS{1-16}:CORR:COLL:METH:SHOR 2
:SENS{1-16}:CORR:COLL:METH:SHOR 3
:SENS{1-16}:CORR:COLL:METH:SHOR 4
:SENS{1-16}:CORR:COLL:SHOR
Select
:SENS{1-16}:CORR:COLL:LOAD
:SENS{1-16}:CORR:COLL:SAVE
:SENS{1-16}:CORR:COLL:CLE
663